GHRS Instrument Science Report 074
STScI Logo

Hubble Space Telescope
GHRS Instrument Science Report 074

Program ID 6898: GHRS Cycle 6:
CRIMP2 - Condensed-Regimen Integrated Monitoring Program for Side 2 of the GHRS

[Previous][Next][Up]


Plan

Purpose: The purpose of this program is to monitor the sensitivity, wavelength and detector calibrations of Side 2.

Description: Four times per year (twice in Cycle 6) the following observations will be done:

DETECTOR CALIBRATIONS:

 1) dark count measurement
 2) focus check at nominal high voltage and tweak current
 3) flat field exposure at center of photocathode
 4) coarse sample mapping function at central y deflection
 5) coarse line mapping function at central x deflection

SENSITIVITY MONITORING

The sensitivity calibration will be monitored with the UV spectrophotometric standard star BD+28D4211. The medium resolution gratings will measure the standard through the LSA and SSA. 5 observations will be made between 1200 and 3000 Å

WAVELENGTH MONITORING

The spectral cal lamp mini-functional test for the first order gratings will be done. It is an internal test which makes measurements of the lamp (SC2). It is used to monitor the carrousel function, y deflections, resolving power, and sensitivity.

Fraction of GO Programs Supported: All GHRS programs using gratings G160M (19%), G200M (7%), and G270M (15%).

Resources:

Observation: Internal-16 orbits (8 orbits for each of the 2 executions. Note: the internal orbits are full orbits). External-6 orbits (3 orbits for each of the 2 executions).

Analysis: 0.24 FTE. The data will be reduced and analyzed by a technical assistants, L. E. Sherbert (0.08 FTE) and A. Gonnella (0.08 FTE) under the supervision of an Instrument Scientist (0.04 FTE per execution for a total of 0.08 FTE).

Special Requirements: One complete set of observations (detector + sensitivity + wavelengths) should be on one SMS. The detector calibrations themselves should be executed within 24 hours. The 3 repeats should occur should occur every 120 +/- 10 days.

Accuracy: Relative sensitivities will be measured to better than 2%. Wavelengths should be good to within 1 diode RMS.

Products: The results of the test will be discussed in an ISR to be published 2 months after completion of the program.