Ideal
Integrating Bolometer
A. Kogut, M. DiPirro, S.H. Moseley, and C. Allen
A bolometer consists of a thermometer with heat capacity C mounted on
an absorbing substrate, weakly coupled to a bath at
temperature T0 via a thermal conductance G (
Figure 1
). Resistive bolometers (including both composite and transition-edge
bolometers) use the temperature dependence of a resistive element to measure
the infrared signal incident on the absorber.
In steady-state operation, the thermometer heats up until the dissipated power matches the power conducted through
the thermal link to the bath,
where Pelec and PIR represent the electrical and infrared power, respectively. Larger temperature excursions — a larger signal — can be achieved by reducing the conductivity G of the bolometer to the bath. To mitigate low-frequency (1/f) noise, though, the infrared signal must be modulated at some frequency w. The synchronous temperature change then becomes
wheret
= C/G is the time constant of the device.
Equation 2 is equivalent to the dc response convolved with a low-pass
filter and illustrates a fundamental limitation of resistive devices: the
conductance G can not be lowered arbitrarily to increase the dc response
without forcing the time constant
t
to unacceptably large values.
For fixed chopping frequency
w
, the sensitivity reaches a plateau at G
»
C
w
beyond which further reductions in G
bring little improvement.
We have developed a new bolometer
concept using a non-dissipative thermometer coupled with a heat switch to
break the link between thermal conductance and performance. Penetration-depth thermometers use inductive coupling to
measure the temperature of a thin superconducting film without dissipating
heat into the system. Temperature
changes in the bolometer thus depend solely
on the integrated infrared power incident on the absorber. The bolometer time constant can then be made arbitrarily
long, allowing the device to function as an ideal (lossless) integrator.
A heat switch periodically increases
the conductance G by a factor of
104 or more during short "reset" intervals.
This forces the bolometer to the bath temperature on time scales short
compared to the intrinsic time constant, eliminating effects of 1/f noise and
thermal drifts on time scales longer than the period between successive
resets.
Figure 1 shows a schematic of the ideal integrating bolometer. It consists of a non-dissipative thermometer on an absorbing substrate, weakly coupled to a thermal bath through a heat switch with conductance ratio g = Gon/Goff . The combination of a non-dissipative sensor with a resettable heat switch breaks the link between conductance and sensitivity that limits resistive bolometers. In the "on'' state the conductance is high and the thermometer quickly cools to the bath temperature. In the "off'' state the conductance is minimal and the thermometer heats as
Provided GDT
«
PIR,
the bolometer responds as an ideal integrator, with temperature change linearly dependent
on the integrated infrared power incident on the absorber.
Penetration-depth thermometers are an attractive technology for the lossless thermometer.
They use the temperature dependence of the
(partial) screening of an external magnetic field by a thin superconducting film to monitor
the temperature of the film without dissipative losses or electrical contacts.
The Meissner effect in a superconductor generates surface screening currents to block an
external magnetic field; the simple case of a bulk superconductor in a parallel magnetic
field B0 leads to an internal field B(x) = B0 exp(-x/
l
)
where l
is the characteristic penetration depth for the field.
Near the superconducting transition, the penetration depth has a strong
temperature dependence
where
Tc is the transition temperature and l
0»
5-20 nm is the penetration depth at T=0.
For film thickness d ~ l0
eal integrator, with temperature change linearly dependent on the integrated infrared power.
The bolometer stays in the linear regime provided
that GD
T«PIR and the integration time Dt is short compared to the time constant t = C / G. Both
of these require small conductance, and hence a long time constant.
We recover device speed and avoid unwanted 1/f contributions by using a
heat switch to link the absorbing substrate to the bath.
The switch effectively produces two time constants: a long time
constant in the "off'' state while integrating the sky signal, and a
short time constant in the "on'' state to reset the device to the bath
temperature. The simplest choice
for a heat switch uses a thin metal strip toggled between its superconducting
and normal states. At
temperatures well below the superconducting transition, all conduction
electrons are paired and the superconductor has small thermal conductivity.
Applying a magnetic field drives the superconductor to its normal
state, greatly increasing the thermal conductivity.
Conduction ratios g > 104 can be achieved using aluminum near
100 mK (Mueller 1978).
A superconducting heat switch
requires a magnetic field large enough to drive the switch to the normal
state, without interfering with the thin film fields nearby.
We achieve this by sandwiching a thin aluminum film between niobium
layers deposited on one support leg of the bolometer. The top and bottom niobium layers form a continuous circuit,
generating an appreciable field within the aluminum strip between them, but
cancelling at distances large compared to the 0.4 mm separation of the sandwich.
A current 5 mA or less through 10 mm
wide niobium strips will generate the 10-2 Tesla critical field
needed to force the aluminum to the normal state, falling below 10-9
Tesla at the penetration depth pickup coil.
Since the niobium legs are deposited on top of each other, the fringe
field at the coil is primarily parallel to the chip surface and will not afect
the perpendicular field used by the bolometer.
Measurements of cross-talk on a test switch agree well with
calculations, providing assurance that cross-talk between the heat switch and
the pickup coil will not limit the performance of the bolometer.
The niobium remains in the superconducting state throughout the entire
reset cycle, adding negligibly to the total thermal conductance.
The inductance of the switch is small (L < 1 nH), allowing rapid
toggling. Figure 2
shows a top view of a prototype bolometer
developed at GSFC. It consists of
a 1 mm ґ 1 mm silicon nitride membrane 0.5 mm
thick, suspended by four silicon nitride legs for low thermal conductance.
A Mo-Au film 60 nm thick on the back side of the membrane provides the
superconducting film, with a 25-turn niobium pickup coil on the front side.
To ensure accurate alignment, the superconducting film and pickup coil
are deposited on opposite sides of the same chip.
Niobium leads for the pickup coil occupy two of the four legs, while
the niobium-aluminum heat switch ``sandwich'' occupies a third.
A thin bismuth film on the membrane provides impedance matching for
greater far-IR absorption. The niobium leads for the heat switch make electrical contact
with the bismuth film; a final heater lead on the fourth leg allows a bias
current to pass through the bismuth film, providing an additional source of
heat for calibration and test purposes. The
prototype bolometer is optimized for operation at 300 mK; greater sensitivity
can be achieved by lowering the temperature to 100 mK.
The sensitivity of a bolometer is described by the noise equivalent power (NEP), the incident infrared power required to produce the minimum detectable signal (see, e.g., Mather 1984 and references therein). The integrating bolometer has several advantages over resistive bolometers. Since it has no dissipative elements, the Johnson noise term vanishes. The heat switch cycle produces an effective response time Dt much smaller than the intrinsic time constant t, resulting in reduced phonon noise. To see this, consider the spectral density of random thermal fluctuations DT within the absorber,
where
The
heat-switched bolometer cannot respond to fluctuations on time scales longer
than the reset period Dt, imposing a low-frequency cutoff on the integral.
With this cutoff, the variance becomes
A non-dissipative,
heat-switched bolometer is more than an order of magnitude more sensitive than
a resistive bolometer.
This can be understood intuitively.
An unswitched bolometer can have only one statistically independent
sample per time interval
t. Activating
the heat switch shortens the time constant by a factor
g
> 104; hence, the sample taken after the heat switch is turned
off again is statistically independent of previous samples.
Using a heat switch allows us to acquire N = t/Dt more samples in a fixed time interval. Since the variance in the mean of N samples scales as 1/N, we
thus reduce the phonon noise by a factor <
span style="mso-char-type: symbol; mso-symbol-font-family: Symbol; font-size: 10.0pt; mso-bidi-font-size: 10.0pt; font-family: Symbol; mso-ascii-font-family: Times; mso-fareast-font-family: MS Mincho; mso-hansi-font-family: Times">жDt/t. An
NEP well below 10-21 W Hz-1/2 appears achievable.
The integrating bolometer has a relatively simple
readout. A persistent current
stored in a superconducting drive coil generates an external field through the
thin-film thermometer. As the
film temperature changes, the inductance of the coil-film system changes.
To conserve flux through the superconducting coil, the coil current
changes and can be measured using a dc SQUID.
The complete measurement cycle has the following steps.
Shortly after the heat switch is turned off, a SQUID measures the
current in the drive coil. The
bolmometer then integrates the infrared signal as the SQUID performs the
initial current measurement on other devices in the array.
After a time Dt short compared to the time constant t, the SQUID re-measures the current to determine the
integrated power absorbed by the bolometer.
We then apply current to the heat switch, driving the aluminum layer to
its normal state and resetting the chip to the bath temperature for the next
cycle. Note that the signal
depends on the difference in current measurements at the start and end of each
cycle, and is insensitive to drifts on time scales long compared to the reset
period. Commercially available SQUIDs have typical flux noise 410-6 F0 Hz-1/2 with 1/f knee at 0.1 Hz, about 800 times the quantum limit. New 2-stage SQUIDS achieve better performance (100 times the quantum limit), at the cost of a higher 1/f knee between 1 and 10 Hz. The amplifier and phonon noise terms scale differently with physical device size. For fixed heat switch period, the phonon noise of an integrating bolometer scales as R-2 from the dependence of the cutoff term on the heat capacity C (assuming film thickness independent of radius). The flux resolution of the thermometer scales as R while the heat capacity scales as R2; hence the amplifier noise ~ C/F scales as R. For a device operated at 50 mK, the NEP reaches a minimum below 10-21 W Hz-1/2 at sensor size 1 mm to 100mm (Fig 3 ), a range easily achieved by photolithographic techniques.
The integrating bolometer has several advantages over resistive devices. The heat switch allows the conductivity to be optimized for sensitivity without sacrificing device speed. The lossless thermometer allows the bolometer to operate as an ideal integrator between resets. The device speed, data rate, and dynamic range depend on the period between heat switch resets and can be changed in real time. Signal information is derived solely by differencing "snapshot" current measurements, without requiring continuous monitoring throughout the integration period. A single readout circuit can thus service multiple devices, reducing the complexity, cost, and power consumption of the readout electronics.